Portable Gamma Survey Meter NHL

Portable Gamma Survey Meter NHL

NHL

NHL
  • High Sensitivity
  • Wide Range Measurement
  • Compact and Lightweight
  • Easy Operation with Single Hand

NHL3

Radiation Measured Gamma (X) rays
Detector 50keV to 6MeV
Measurement Range Dose Rate: 0.01μSv/h to 99.99mSv/h
Accumulated Dose: 0.001μSv to 99.99mSv
Energy Response ≤ ± 30% (60keV to 1.5MeV,137Cs reference)
Accuracy ≤ ± 10%(1μSv/h to 99.99mSv/h)
Angular Response ≤ ± 25%
(Horizontal/ Vertical ± 90° :137Cs except Battery direction)
Time Constant Setting Standard deviation fixed mode or Time constant fixed mode(switchable)
Time Constant : 3level setting
(3seconds, 10seconds, 30seconds)
Display LCD
Indication Data Current dose rate, Maximum dose rate,
Accumulated dose, Alarm setting values,
Battery indication and Time indication
Sound Function Counting and Alarm sound
Power Supply Standard AA Batteries (LR6) x 2
Alkaline or Nickel-metal-hydride
(USB supply, recharging possible)
Battery Lifetime* ≥ 72 hours
Communicating Functions USB
Operating Temperature -20℃ to 50℃ (-4°F to 122°F)
Operating Humidity ≦ 95%RH (Non-condensing)
Degree of Protection(IP Code) IP54
Dimensions 60(W) x 27(D) x 120(H) mm approx.
( 2.4(W) x 1.1(D) x 4.7(H) in approx.)
Weight Less than 230g / 0.51lb

*Continuous operation, when counting sound OFF, communication f unction OFF
1R=100Sv

NHL4

Radiation Measured Gamma (X) rays
Detector 50keV to 6MeV
Measurement Range Dose Rate:
0.001μSv/h to 99.99mSv/h 0.1μR/h to 9.999R/h
Accumulated Dose:
0.001μSv to 99.99mSv 0.1μR to 9.999R
Energy Response ≤ ± 30% (60keV to 1.5MeV,137Cs reference)
Accuracy ≤ ± 10%
(0.1μSv/h to 99.99mSv/h, 10μR/h to 9.999R/h)
Angular Response ≤ ±25%
(Horizontal/ Vertical ± 90° :137Cs except Battery direction)
Time Constant Setting Standard deviation fixed mode or Time constant fixed mode(switchable)
Time Constant : 3level setting
(3seconds, 10seconds, 30seconds)
Display LCD
Indication Data Current dose rate, Maximum dose rate,
Accumulated dose, Alarm setting values,
Battery indication and Time indication
Sound Function Counting and Alarm sound
Power Supply Standard AA Batteries (LR6) x 2
Alkaline or Nickel-metal-hydride
(USB supply, recharging possible)
Battery Lifetime* ≥ 48hours
Communicating Functions USB, GPS
Operating Temperature -20℃ to 50℃ (-4°F to 122°F)
Operating Humidity ≤ 95%RH (Non-condensing)
Degree of Protection(IP Code) IP54
Dimensions 60(W) x 27(D) x 120(H) mm approx.
(2.4(W) x 1.1(D) x 4.7(H) in approx.)
Weight Less than 260g / 0.57lb

*Continuous operation, when counting sound OFF, communication f unction OFF
1R=100Sv

Top of Page